Advantages of the Cough Assistant RespiClear:
- User-friendly operation and interface.
- Fast switching times.
- Patented valve technology.
- Adjustable oszillation levels
- Constant flow limitation.
- Space-saving slim design.
- Can also be used when lying down.
- Comfortable setting on the 7 inch touchscreen.
- Easy cleanable surfaces.
- Up to 5 presets can be saved.
- Control via foot pedal optionally available.
- Carrying bag designed for mobile use.
- Very low noise .
- Optimum protection of the device and the breaching air through a cartridge filter class M5
Characteristics:
Clear display with a structured menu
Adjustable parameters:
- Manual mode
- AUTO / AUTO with trigger
- PC AirStacking
- NIV
Up to 5 previously configured presets
- Options for copying
- You can transfer your data from one program to another.
- Deactivating the options.
Various therapy options and additional functions:
- Vibrations from 4 to 15 Hz, during exhaling or / and / inhaling.
- Automatic maximum amplitude generation.
- Flow restriction from 100% to 80% or 60%.
- Interruption of the I / E for sensitive patients, even those with Dysphagia.
- Gentle increase of ramp pressure for better patient compliance.
Review of patient information and Peak Cough Flow report.
NIV Mode
- Limited ventilation time (max. 30 min.)
- Suitable for preparing extremely exausted patients before and / or after ventilation, for better therapy effect and quicker compliance with the machine.
Techical data
Product class acording to EC directive 93/42/EEC |
IIa |
Electrical safety - Protection class - Protection level - Protection rating - Complies with standard |
II (totally insulated B IP 21 DIN EN 60601-1:2013 |
Electrical connection - Alternating current - Direct current |
100 - 240 V AC automatic, 50 - 60 Hz Tolerance -15% to +10% 24 V DC / 5 A Tolerance in each case -15% to +25% |
Power - Maximum power |
215 VA / 135 W |
Ambient conditions -Operation Temperature Air humidity Air pressure |
+5°C to +40°C 15% RH to 95% RH 500 hPa to 1200 hPa
|
Operation | Continous operation |
Operating range |
Manual mode, AUTO, AirStacking: +70hPa to -70hPa NIV: 4hPa to 30hPa The maximum pressure of <60hPa will not be exceeded, even in the event of the first fault listed. |
Sound pressure level alarm messages | min. 70dB(A) |
System resistance with air flow of 60 l/min at patient connection opening | RespiClear with hose system and bacterial filter 0,30 kPa s / l |
Minimum stable pressure limit (PLSmin) | > 0 hPa (minimum pressure in event of fault) |
Maximum stable pressure limit (PLSmax) | < 60 hPa (maximum pressure in event of fault) |
Dimensions (WxHxD) | 35 x 31 x 15 mm |
Weight | 4,9 kg |
Air filtering -filter life |
12 months |
File | Info |
---|---|
iHope RS Ventilator Datasheet | [PDF, 679.82 KB] |